Comparison of Lead and Cerrobend Blocks for Incident Photon Flux of 6 and 15 MV X‐rays

AUTHORS

Ahmad Iftikhar 1 , * , M Wazir 2 , MB Kakakhail 3 , A Sbilal 3 , H Amjad 2 , A Khwaja 1 , A Khushnaseeb 1

1 Center for Nuclear Medicine And Radiotherapy (CENAR), BMC Complex, Quetta, Pakistan

2 Institute of Nuclear Medicine, Oncology and Radiotherapy, Abbotabad, Pakistan

3 Pakistan Institute of Engineering and Applied Sciences (PIEAS) Islamabad,, Pakistan

How to Cite: Iftikhar A, Wazir M, Kakakhail M, Sbilal A, Amjad H, et al. Comparison of Lead and Cerrobend Blocks for Incident Photon Flux of 6 and 15 MV X‐rays, Int J Cancer Manag. 2011 ; 4(1):e80721.

ARTICLE INFORMATION

International Journal of Cancer Management: 4 (1); e80721
Published Online: March 30, 2011
Article Type: Research Article
Received: September 01, 2010
Accepted: December 04, 2010

Crossmark

CHEKING

READ FULL TEXT
Abstract

Background: During radiotherapy treatment, critical organs are shielded using lead and cerrobend blocks. The objective of this study is to compare the effects of lead and cerrobend shielding blocks on incident photon beam.

Methods: Collimator scatter factors were measured for open square fields (3 x 3 cm to 40 x 40 cm) defined by collimator jaws and for fields blocked down to smaller asymmetric fields by using five different Lead and Cerrobend blocks for 6 and 15 MV photon beams from a Varian Clinac 2100C accelerator. The measurements in air were performed using Farmer type ionization chamber fitted with acrylic build up caps.

Results: The Block Tray Factor (BTF) increased with field size for both 6 and 15 MV photon beams. In case of Lead blocks, the extreme variations in BTF for 6 MV photon beam are 0.70 %, 0.84 %, 0.56%, 0.80 % and 1.15 %. Similarly, for 15 MV the maximum variations for Lead blocks are 0.46 %, 0.60 %, 0.83 %, 0.88 % and 1.10 % respectively. No significant difference has been observed in the BTF of Cerrobend blocks for 6 and 15 MV photon beams.

Discussion: The dose received by a point in air apparently shielded by lead blocks has three main contributions: 1. Due to primary photon beam transmitted through the block, 2. Due to scattered photons, 3. Due to contamination electrons. These three factors collectively cause the increase in BTF with increasing field size, energy, and decreasing block size.

Conclusions: The effect of shielding on the beam output increases with field size, beam energy and shield size. This increase follows almost the same pattern for both lead and cerrobend shielding blocks. Therefore shield factors for all field sizes, beam energies and shield sizes should be determined separately for precise patient dose delivery.

Keywords

Cerrobend shielding blocks Output factor Build up caps

© 2011, Author(s). This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/) which permits copy and redistribute the material just in noncommercial usages, provided the original work is properly cited.